Laryngology · Other devices

Pharyngeal Tube

Flexible biocompatible silicone tube that collects saliva in the oropharynx and channels it to the esophagus, preventing tissue maceration and aspiration into the airway. Part of the Stening® laryngology line.

Product code: TF12
Biocompatible silicone pharyngeal tube Stening® (TF)
Pharyngeal tube or pharyngeal prosthesis of silicone Stening®
Stening® pharyngeal tube proximal cup view
Application of the pharyngeal tube in esophageal stenosis
Application of the pharyngeal tube in esophageal carcinoma with fistula
Laryngology · Other devices

Pharyngeal Tube

Code TF12

The Pharyngeal Tube —also called pharyngeal prosthesis or oropharyngeal tube— is made of flexible silicone and is useful in a variety of situations. Its apical end widens progressively: thus the anterior edge keeps contact with the base of the tongue, while the posterior edge rests on the pharyngeal wall.

In this way it collects saliva in the oropharynx and channels it to the esophagus, preventing tissue maceration and avoiding the aspiration of saliva into the airway. A nasoenteral feeding tube can be introduced through the Pharyngeal Tube.

Material
Flexible biocompatible silicone
Line
Laryngology / Tracheostomy
Presentations
Single size (TF12)
Technical specifications

Product information

Review the indications, dimensions, how to use, care and warnings of the Pharyngeal Tube.

Clinical indications

The Pharyngeal Tube is useful in a variety of situations that require channeling saliva to the esophagus and protecting the airway.

  • Fistulas secondary to laryngectomy, radiotherapy, neoplastic conditions and caustic ingestion
  • Traumatic orocutaneous or pharyngocutaneous fistulas
  • Head and neck oncological surgery
  • Esophageal stenosis
  • Esophageal carcinoma

Available dimensions

The Pharyngeal Tube is offered in a standard size (TF12) and can be cut to the length deemed appropriate for each case.

Dimensional diagram of the Pharyngeal Tube TF12
CodeTube ØProximal cupTotal length
TF121237185

Measurements expressed in millimeters. The tube can be trimmed to the appropriate length for each case. For specific sizing inquiries, contact us at (+54) 11 4553-5070 or (+54) 11 4551-2333.

Insertion technique

The insertion of the Pharyngeal Tube must be performed under general anesthesia. It can be cut to the length deemed appropriate for the case.

A laryngoscope will be used to gain comfortable access to the larynx and adequate vision. With the help of a hypopharyngoscope and long forceps, grasp the tube by its distal end to guide it toward the esophagus, until the proximal cup is at the laryngeal level. Prior esophageal dilation may be necessary. After its insertion it can be fixed with a percutaneous suture.

Care

Periodic medical check-up.

Warning for use

Important

The device must not be reused, as this could cause cross-contamination.

Also known as: pharyngeal prosthesis · oropharyngeal tube · salivary bypass tube · TF

Need more information